Establishment of specifcations of the multiplex real-time PCR kit (NIHE-RealtimePCR hrHPV) for the detection of high-risk HPV genotypes

Các tác giả

  • Chu Thi Ngoc Mai National Institute of Hygiene and Epidemiology, Hanoi
  • Nguyen Van Chien Suran Medical and Scientifc Solution, Hanoi
  • Luu Thi Dung National Institute for Control of Vaccines and Biologicals, Hanoi
  • Tran Thi Nguyet Lan National Institute of Hygiene and Epidemiology, Hanoi
  • Le Thi Khanh Ly National Institute of Hygiene and Epidemiology, Hanoi
  • Dang Duc Anh National Institute of Hygiene and Epidemiology, Hanoi
  • Nguyen Van Trang National Institute of Hygiene and Epidemiology, Hanoi

DOI:

https://doi.org/10.51403/0868-2836/2021/333

Từ khóa:

Human papillomavirus (HPV), cervical cancer, real-time PCR

Tóm tắt

High-risk human papillomavirus (HPV) genotypes are the leading causes of anogenital cancer, including cervical cancer. Screening these genotypes will potentially alleviate the disease burden that HPV causes. Therefore, NIHERealtime PCR hrHPV was developed based on the principle of multiplex real-time PCR. The objective of this study is to establish the specifcations of this assay in detecting and identifying 14 high-risk genotypes of HPV. The specifcations of this assay including cross-reactivity, sensitivity and specifcity, the limit of detection and stability, were evaluated with clinical samples, HPV positive control code 06/202 and 06/206 (National Institute for Biological Standards and Control) and in-house plasmids. Four different real-time systems and two commercial assays were applied for comparison. No cross detection of bacteria and viruses related to other genital-anal was observed. The sensitivity and the specifcity of this kit both reached 100% for HPV-18, 92.3% and 100% for HPV-16, and ranged from 90.0% - 100% for the other 12 high-risk genotypes (n = 92) with the detection limit of 5 copies/µl for HPV-16 and HPV-18, and 10 copies/µl for the other 12 high-risk genotypes HPV. This test was stable for up to 12 months under -20oC condition. Equivalent performance was demonstrated on all four real-time systems. Finally, the product was highly comparable with the Sacace kit, Italy (HPV Genotypes 14 Real-™ Quant, V67-100FRT), evaluated by the World Health Organization. The NIHE kit has the potential to be used in the early detection of HPV infection to improve control, prevention and treatment of HPV-related anogenital diseases.

Tải xuống

Dữ liệu tải xuống chưa có sẵn.

Đã Xuất bản

18-06-2021

Cách trích dẫn

Mai, C. T. N. ., Chien, N. V. ., Dung, L. T. ., Lan, T. T. N., Ly, L. T. K. ., Anh, D. D. ., & Trang, N. V. . (2021). Establishment of specifcations of the multiplex real-time PCR kit (NIHE-RealtimePCR hrHPV) for the detection of high-risk HPV genotypes. Tạp Chí Y học Dự phòng, 31(4), 28–36. https://doi.org/10.51403/0868-2836/2021/333

Số

Chuyên mục

Bài báo nghiên cứu gốc

Các bài báo được đọc nhiều nhất của cùng tác giả

1 2 > >> 

Các bài báo tương tự

Bạn cũng có thể bắt đầu một tìm kiếm tương tự nâng cao cho bài báo này.